Continuous Glucose Monitoring for Adults: Position Statement
v3.1 March 2026
1. Introduction
People with diabetes on insulin need to regularly monitor
their blood glucose levels to judge the amount of insulin they need to inject,
to keep their blood glucose within an acceptable range and reduce the risk of
short- and long-term complications.
Historically, there were three main types of blood glucose
monitoring: self-monitored blood glucose testing (using skin prick tests),
real-time continuous glucose monitoring (rtCGM) and intermittently scanned
continuous glucose monitoring (isCGM). isCGM is also sometimes known as
‘flash’.
CGM devices can be divided into "low-cost” and
"high-cost”. Low-cost CGM are those which cost <£1000 and are usually
available to prescribe via a FP10. Formulary choices include FreeStyle Libre 2
plus, Dexcom One+ and CareSens Air sensor. These devices have sufficient
functionality to meet the needs of most patients requiring CGM (see appendix
1).
Higher cost CGM devices >£1000 on the formulary include
Freestyle Libre 3, Dexcom G6 & G7, Medtronic 3 & 4. These have
additional functionality which is appropriate for selected groups of patients
(see appendix 2). Freestyle Libre 3 can now be upgraded to the FreeStyle Libre
3 plus which offers even more functionality.
This guidance update adds eligibility criteria for each of
the different low and high cost CGM devices available locally, to ensure that
the most cost-effective, clinically appropriate device is selected. This is
essential to support any future widening of access to CGM for adults.
2. Content
Eligibility for CGM
The following cohorts are eligible for CGM:
- Adults
with type 1 diabetes who are willing to commit to using CGM at
least 70% of the time and to calibrate it as needed, and who have any of
the following despite optimised use of insulin therapy and conventional
blood glucose monitoring (As per the
previous National Institute for Health and Care Excellence (NICE) guidance
NG17 and criteria from NHS England (NHSE) 2019 Flash guidance):
- More
than 1 episode a year of severe hypoglycaemia with no obviously
preventable precipitating cause.
- Impaired
or complete loss of awareness of hypoglycaemia.
- Frequent
(more than 2 episodes a week) asymptomatic hypoglycaemia that is causing
problems with daily activities.
- Extreme
fear of hypoglycaemia.
- Are
clinically indicated as requiring intensive monitoring >8 times daily,
as demonstrated on a meter download/review over the past 3 months
- Are
unable to self-monitor due to disability
- The
specialist diabetes MDT determines have occupational or psychosocial
circumstances that warrant a 6-month trial of CGM
- Are
living with a learning disability and it is recorded on their GP Learning
Disability register
- Are
pregnant (12 months in total inclusive of post-delivery period)
- Adults
with insulin treated type 2 diabetes who are living with a
learning disability and it is recorded on their GP Learning Disability
register (As per NHSE 2019 Flash guidance,)
- People
with any form of diabetes on haemodialysis and on insulin
treatment who are clinically indicated as requiring intensive monitoring
>8 times daily, as demonstrated on a meter download/review over the
past 3 months (As per NHSE 2019 Flash guidance,)
- People
with diabetes associated with cystic fibrosis on insulin treatment (As
per NHSE 2019 Flash guidance,)
A summary on process for initiation and supply of low cost
CGM has been added as appendix 3.
The Dexcom One+, FreeStyle Libre 2 plus and
CareSens Air have sufficient functionality to meet the needs of most patients
requiring CGM. They all have customisable hypoglycaemia alerts, supporting
those with impaired hypoglycaemia awareness. They are both able to do repeat
hypoglycaemia alerts (e.g. if hypoglycaemia continues or reoccurs).
CareSens Air is currently the most affordable low cost CGM
option and has been added to the formulary. The CareSens Air has a warm up time
of only 30 minutes, allowing patients to start tracking their readings much
sooner. There is an all-in-one applicator for simple use, and it is designed
for up to 15 days of wear, provides real-time glucose readings every 5 minutes
via bluetooth. These readings can be securely shared with unlimited followers.
The Freestyle Libre 2 plus is an upgrade to the Freestyle Libre 2. It continues
to allow data sharing. The sensors have 15 day wear (opposed to 14) and can
link with specific insulin pumps to form a hybrid closed loop system.
Initiation of CGM is via specialist diabetes teams only.
Applications for low-cost CGM should be made by the specialist teams via the
Blueteq drug management system. The low-cost CGM devices, following initiation
by the specialists, can be prescribed via FP10 in primary care.
See Appendix 1 for a
comparison document for formulary CGM devices that are available on FP10.
- Switching
between low-cost CGM manufacturers
Patients currently receiving NHS-funded low-cost CGM who
feel that an alternative manufacturer’s system is better suited to managing
their diabetes should be advised to discuss this with their diabetes specialist
at their next routine appointment. Specialists may switch patients to an
alternative manufacturer’s device where considered suitable. Primary care
clinicians should only switch the patient’s prescription to an alternative
manufacturer’s device when advised to do so by the patient’s diabetes specialist.
- Purchasing
CGM privately
CGM prices in appendix 1 relate to prices to the NHS.
Patients may wish to purchase CGM from a pharmacy on a private basis. The
private purchase price of CGM will vary between different pharmacies and may
include VAT. Where patients have diabetes, they might not have to pay VAT. VAT
exemption needs to be discussed with the relevant pharmacy where the patient is
purchasing the CGM and may be only claimed at the time of purchase. The
diabetes UK website contains information that patients may find helpful in relation
to VAT
Exemption for People with Diabetes.
Eligibility for high-cost CGM
Some patients will require additional functionality that is
only available on high-cost CGM devices. To ensure the most cost-effective,
clinically appropriate device is chosen, patients requiring a high-cost CGM
device must meet both the eligibility criteria above for CGM and the criteria
for the relevant high-cost CGM device.
After funding for high-cost CGM has been approved, if a
patient’s circumstances change such that they are no longer eligible for a
high-cost device, they should be switched to a low-cost device at the earliest
opportunity, rather than at the end of the approved funding period.
The following criteria are based on the additional
functionality and cost of each device (see appendix2).
FreeStyle Libre 3 plus
The FreeStyle Libre 3 plus is the least
expensive of the formulary high-cost devices. It can be used as a standalone
CGM with optional high/low glucose alarms and a mandatory urgent low alarm and
also has data sharing capability It can be linked with a compatible insulin
pump to create a hybrid closed loop system.
Criteria: One of the following must apply
- For use in children aged 2+ who have tried low cost FSL2/
FSL2 plus and Dexcom One / Dexcom One Plus, but are unable to wear the sensors
due to the size and require a smaller sensor. Evidence must be provided.
- Eligible for a hybrid closed loop system and where
currently using or about to start a compatible YPSOmed pump with CamAPS.
Initiation of FreeStyle Libre 3 plus is via specialist
diabetes teams only. Applications should be made by the specialist teams via
the Blueteq drug management system. An initial supply of 28 days will be
supplied by the specialist teams. Primary care can be requested to take up
on-going prescribing after initiation. The devices can be prescribed via FP10.
Information for Primary Care
- Upgrading
from Freestyle Libre 3 to Freestyle Libre 3 Plus
All existing users of Freestyle Libre 3 will require
upgrade to Freestyle Libre 3 Plus. Secondary care is responsible for
identifying and upgrading patients.
FreeStyle Libre 3 plus
The Freestyle Libre 3 plus is an upgrade to the Freestyle
Libre 3. The Libre 3 app and the libre 3 reader work with the new sensor, no
app updates or change of reader are required. It continues to allow data
sharing. The sensors have 15 day wear (opposed to 14) and has an expanded age
indication of 2 years and older.
Dexcom G7
In addition to the functionality of the low-cost CGM
devices and Free Style Libre 3 plus, the Dexcom G7 has predictive alerts, which
warn the wearer that they will be hypoglycaemic soon. These allow more time to
act to prevent or self-treat hypoglycaemia. Like the Dexcom ONE plus, the G7 is
able to do repeat hypoglycaemia alerts (e.g. if hypoglycaemia continues after
the initial alert or reoccurs within 15 mins), and is also able to share
readings. Dexcom G7 a cost-effective option where predictive alerts are
required.
Criteria:
- Have tried low cost real-time CGM
AND
- Require predictive alerts due to at least one of the following whilst on low
cost real-time CGM:
- More
than 1 episode in a year of severe hypoglycaemia (i.e. requiring 3rd
party assistance) with no obviously preventable precipitating cause
OR - Frequent
(more than 2 episodes a week) asymptomatic hypoglycaemia that is causing
problems with daily activities.
OR - Extreme
fear of hypoglycaemia that persists despite 6 months of low cost
real-time CGM.
Dexcom G6
Used
with a compatible pump, the G6 can form a hybrid closed loop (HCL) or
predictive low-glucose suspend (PLGS) sensor-augmented pump
Criteria:
- Using an insulin pump which requires the G6 to create a
hybrid closed loop (HCL) or predictive low-glucose suspend (PLGS)
sensor-augmented pump.
OR
- Currently eligible for an insulin pump and intend to start pump therapy with
a compatible pump within the next 3-6 months.
Medtronic 3 and Medtronic 4
The Medtronic 3 and 4 are not standalone CGMs and can only be used with the
Medtronic pump as part of a low-glucose suspend (LGS) sensor-augmented pump,
predictive low-glucose suspend (PLGS) sensor-augmented pump or hybrid closed
loop (HCL) system.
Criteria:
- Currently using or about to start a compatible Medtronic
pump.
Other indications for high cost CGM
Sleeping through hypoglycaemia alarms is not an indication to switch to a
higher-cost CGM. This is because manufacturer specifications do not indicate
that higher cost CGM have louder alarms. Differences in smartphone devices may
have more impact on alarm volume than the CGM device (and its corresponding
app). Patients should be supported to ensure that their smartphone is optimally
configured.
Funding for higher-cost CGM devices will be considered due
to skin reactions to sensor adhesives where there is:
- Significant, documented, contact dermatitis in response
to sensor adhesives that cannot be managed otherwise (e.g. through use of
barrier products)
OR
- Documented allergic reaction to sensor adhesive.
The lowest cost, clinically suitable device should be
selected.
Indications outside of the criteria listed in this guidance
will be considered on an individual basis via the IFR route, with a review of
the guidance criteria if relevant additional cohorts are identified.
3. Monitoring
Diabetes technology is a rapidly changing area. This
guidance will be closely monitored and reviewed in response to significant
changes, such as the publication of the NICE TA943 on hybrid closed loops (Dec
2023)⁶ and subsequent ongoing work that is being undertaken by NHSE and NHS
supply chain on products that will meet the cost-effectiveness thresholds set
in NICE TA943.
Data to be monitored for all patients on CGM, and all
patients eligible for CGM who have not taken it up:
- HbA1c
- Time
in range
- Number
of blood glucose testing strips (BGTS) used
- Hospital
admissions for diabetes emergencies such as diabetic ketoacidosis (DKA) or
severe hypoglycaemia.
- Episodes
of severe hypoglycaemia which did not result in hospital admission.
We would expect to see an improvement in these indicators
in patients on CGM. We would expect to see an improvement in HbA1c, an increase
in time spent in range, and a reduction of BGTS used, hospital admissions and
episodes of severe hypoglycaemia.
As per the 2022 NICE guidance regarding addressing
inequalities in CGM access and uptake, other actions to be taken include:
- Monitoring
who is using CGM – age, sex, ethnicity, deprivation
- Identifying
groups who are eligible but have a lower uptake
- Making
plans to engage with these groups to encourage them to consider CGM. (1)
(2)
There should also be some qualitative data monitoring to
capture insights on improvements in quality of life. The data collected using
the above indicators and through qualitative methods will be used to evaluate
the CGM guidance and to inform decisions and next steps at future guidance
review stages.
4. References
NHSE 2019 Flash guidance
- National
Institute for Health and Care Excellence. NG18 Diabetes
(type 1 and type 2) in children and young people: diagnosis and
management. 2022.
- National
Institute for Health and Care Excellence. NG17
Type 1 diabetes in adults: diagnosis and management. 2022.
- National
Institute for Health and Care Excellence. NG28Type 2
diabetes in adults: management. 2022.
- Digital,
NHS. Quality and Outcomes Framework, 2020-21. 30
September 2021.
- NHS
Digital. National Diabetes Audit, 2020-21
Quarterly Report. 2021.
- National Institute for Health and Care
Excellence. TA943 (2023) Hybrid closed loop systems for managing blood
glucose levels in type 1 diabetes
5. Associated documentation
- Continuous
Glucose Monitoring for paediatrics – Position statement v3.1 (March 2026)
- Insulin Pumps – Position statement v3.1
(March 2026 )
- Hybrid
Closed Loop Systems for Adults – Local implementation plan v2.2 (March
2026
- Hybrid
Closed Loop Systems for Children and Young People Position Statement v1.0
(July 2024)
Appendix 1: Key features of standalone CGM
available on FP10
Appendix 2: Indicative Cost and Functionality
of locally available CGM devices (at time of publishing).
Appendix 3: Summary of process for initiation
and supply of low cost CGM
Appendix 4: Frequently Asked Questions for low
cost CGM
1. Replacing faulty sensors/transmitters or
sensors that fall off before they are due to be changed
If the sensor or transmitter is suspected to be defective
or if the sensor falls off before it is due to be changed, patients should
contact the manufacturer as soon as possible to obtain a replacement (see
contact details below). Patients will need to keep the defective
sensor/transmitter and follow the instructions given by the
representative. Specialists/GPs should not issue additional
prescriptions to replace defective sensors/transmitters or sensors that have
fallen off.
Dexcom Technical Support Line
Tel: 0800 031 5763
Details of office opening hours can be found at: https://www.dexcom.com/en-gb/contact-us-direct
Abbott Customer Careline
Tel: 0800 170 1177
Details of office opening hours can be found at: https://freestylediabetes.co.uk/contact-us
Medtronic Product/technical support
Tel: 01923205167
Support
can be provided through their app.
2. Quantities of blood glucose test
strips/lancets
Patients who use CGM still need to take blood glucose
measurements, but less often. Blood glucose monitoring is
required to check the accuracy of their CGM device and as a backup when blood
glucose levels are changing quickly, or the device stops working. When CGM is
commenced, prescribed quantities of blood glucose test strips/lancets should
reduce. As per NHS England guidance, patients using a CGM are expected to be
provided with 50 blood glucose test strips per month (one box per month),
subject to the clinical prescriber’s discretion. Specialists should advise
primary care clinicians of the change in quantity of blood glucose strips and
primary care should amend the repeat prescriptions (reduction in number of
strips).
For further information, please see Guidance
for Self-monitoring of Blood Glucose (SMBG) in patients with Diabetes Mellitus.
3. Using CGM to monitor blood glucose levels
for the purpose of driving
The Driver and Vehicle Licensing Agency (DVLA) has
published guidance relating to the use of CGM to monitor blood glucose levels
for the purpose of driving – click here to
access the guidance
Version control
V3.1 March 2026: Following review by HWE ICB Diabetes
Technology Group document type designated as Position statement and removal of
references to policy.